WIND SYSTEM CALIBRATION
RECOMMENDED CALIBRATION INTERVAL,
PROCEDURE, AND TEST EQUIPMENT
MODEL 18860-90
INCLUDES INSTRUCTIONS FOR THE FOLLOWING
MODEL 18802/18811 ANEMOMETER DRIVE
MODEL 18112/18212 VANE ANGLE FIXTURE
MODEL 18310/18312 TORQUE DISC
MODEL 18331 VANE TORQUE GAUGE
Przeglądanie stron 0
1 2 3 4 5 6 ... 15 16

Podsumowanie treści

Strona 1 - METEOROLOGICAL INSTRUMENTS

METEOROLOGICAL INSTRUMENTSINSTRUCTIONSR.M. YOUNG COMPANY 2801 AERO PARK DRIVE, TRAVERSE CITY, MICHIGAN 49686, USATEL: (231) 946-3980 FAX: (23

Strona 2

Page 918860-90R. M. YOUNG COMPANYTYPICAL TORQUE VALUESFor Checking Anemometer Bearing and Transducer ConditionNOTES:1. New instrument torque and th

Strona 5

Page 1218860-90R. M. YOUNG COMPANYTYPICAL TORQUE VALUESFor Checking Anemometer Bearing and Transducer ConditionNOTES:1. New instrument torque and t

Strona 7 - WARRANTY

Page 1418860-90

Strona 8

Page 1518860-90

Strona 9

Page 118860-90R. M. YOUNG COMPANYWIND SYSTEM CALIBRATIONRecommended Calibration Interval, Procedure, and Test EquipmentRECOMMENDED CALIBRATION INTERVA

Strona 10 - R. M. YOUNG COMPANY

Page 218860-90PROCEDUREIn calm weather blow gently on propeller. Watch for obvious high torque or irregular rotation.Remove propeller or cupwheel. Dri

Strona 11 - 18860-90

Page 318860-90LABORATORY CHECKSensors and signal conditioning modules are removed from eld site to calibration lab equipped with calibration xtures

Strona 12

Page 418860-90

Strona 13

Page 518860-90MODEL 18802/18811SELECTABLE SPEEDANEMOMETER DRIVESPECIFICATIONSOperating Range: 200 - 15,000 RPM (18802) 20.0 - 990.0 RPM (18811

Strona 14

Page 618860-90WARRANTYThis product is warranted to be free of defects in materials and construction for a period of 12 months from date of initial pur