Page 3
46203-90
MINIMUM / MAXIMUM
TheTempTrackerstoresminimumandmaximumvaluesforboth
displays.Toviewminimumandmaximumvalues,depressthe
SET/ RESET key momentarily. Each time the key is depressed, the
displayswitchesmodes;rsttoMIN then MAXthenbacktoCur-
rent values. To reset all values, press and hold the SET / RESET
keyuntiltheunitbeeps.
TheTempTrackermaybesettoautomaticallycyclethroughMIN,
MAX and Current values. This feature is selected during SETUP.
Minimumandmaximumvaluesresetifpowerisinterrupted.
REMOTE DISPLAYS
TheTempTrackermaybeusedasaremoteorslavedisplayby
selecting InP SEr during SETUP. Remote displays are connected
totheMaindisplayusingtheRS-485terminals.Connectlike
terminals as shown in the wiring diagram. Remote displays show
exactlythesameinformationastheMaindisplay.MIN/MAXand
SETUP functions are controlled at the Main unit only. Brightness is
adjusted at each display independently.
VOLTAGE OUTPUTS
TheTempTrackerofferscalibratedvoltageoutputsforbothchan-
nels. This feature allows the use of recorders and other devices.
Fullscalevoltageforeachchannelis5.00VDC.
ERROR MESSAGES
The Temp Tracker will detect and indicate certain errors. Once cor-
rected, the error indication disappears.
Display Error
LEFT RIGHT
- - - - - - 4-20 mA signal from sensor is not present or is
outsideacceptablerange.
SEr Err Unitissetuptoreceiveserialinputsignal,but
nodataisbeingreceived.Verifythatmainunit
is operating.Checkcableconnections.
WARRANTY
TheTempTrackeriswarrantedtobefreeofdefectsinmaterialsand
workmanship for a period of 12 months from the date of purchase.
This warranty is limited to repair or replacement of defective unit.
SPECIFICATIONS
• Size 144x144x36mm(5.65x5.65x1.4in)
• PanelCutout 138x138mm(5.43x5.43in)
• SensorType YOUNGModel41382LC2RH/T
YOUNG Model 41342LC T only
• InputSignal T 4-20mA=-50to+50°C
RH 4-20mA=0to100%
• SerialI/O RS-485,binarydatablock,
16blockspersecond
• Accuracy Temperature±0.1°C(0.2°F)
RelativeHumidity±0.1%
DewPoint±0.5°C(0.9°F)
(These accuracies are for the Temp Tracker
only and are independent of sensor accuracy.)
• VoltageOutput 0.00to5.00VDC
Scale Temperature and Dew Point:
-50to50°C,-50to150°F
RelativeHumidity:0to100%
• Alarmrelays Normallyopen,non-latching,
5Aresistive,2Ainductive@24VACor
30 VDC
• Power
requirement 12to30VDC,4.5watts
• Weight 0.75lbs(0.34kg)
• Accessories Model06280RackMountingPanel
CE COMPLIANCE
ThisproducthasbeentestedandshowntocomplywithEuropeanCE
requirementsfortheEMCDirective(seeDeclarationofConformity
below).Shieldedcablemustbeused.
Declaration of Conformity
R. M. Young Company
2801 Aero Park Drive
Traverse City, MI 49686 USA
Model 46203 TEMP TRACKER
TheundersignedherebydeclaresonbehalfofR.M.
YoungCompanythattheabove-referencedproduct,to
which this declaration relates, is in conformity with the
provisions of:
CouncilDirective2004/108/EC(December15,2004)
onElectromagneticCompatibility
David Poinsett
R&DManager
Komentarze do niniejszej Instrukcji